MINDRAY DATASCOPE ACCUTORR PLUS

Category:

Description

Mindary Datascope Accutorr Plus Features

The Mindray Datascope Accutorr Plus is available in 3 models. The Accutorr Plus NIBP, the Accutorr Plus with NIBP with Trend Screen, and the Accutorr Plus with NIBP, Trend Screen and SpO2. The Accutorr Plus with Trend Screen incorporates an LCD to view stored measurements and access a menu system which is used to set the alarm volume and the SpO2 beep volume. It can store up to 100 measurements in its memory.

  • Non-invasive blood pressure (NIBP).
  • Pulse Rate.
  • SpO2.
  • Alarms.
  • Interval Mode.
  • Large LED Displays.
  • Trend Memory up to 100 measurements.

Mindary Datascope Accutorr Plus Specifications

Dimensions

  • Height: 10 in (26cm)
  • Width: 7.5 in (19 cm)
  • Depth: 6.75 in (17 cm)
  • Weight: <10 lbs (4.5 kg)

Systolic Pressure

  • Number of Digits: 3
  • Accuracy: Mean error less than ±5 mmHg, Standard deviation less than ±8 mmHg
  • Range: Adult Mode: 55 to 260 mmHg; Pediatric Mode: 55 to 160 mmHg

Diastolic Pressure

  • Number of Digits: 3
  • Accuracy: Mean error less than ±5 mmHg, Standard deviation less than ±8 mmHg
  • Range: Adult Mode: 30 to 200 mmHg; Pediatric Mode: 30 to 150 mmHg

NIBP Measurement Cycle

  • Less than 30 seconds average at 80 BPM with 180 mmHg pump up pressure, without retries, motion artifact or arrhythmia with standard adult cuff on health individual. Cycle time is affected by arm size and wrapping technique.

Pulse Rate

  • Range: 35-245 BPM for Adult and Pediatric
  • Display Resolution: 1 BPM
  • Accuracy: ±3 BPM or ±3%, whichever is greater

Maximum Cuff Pressure

  • Two means of limiting cuff pressure are provided; a hardware overpressure monitor which limits the pressure to 330mmHg for adults and 220mmHg for pediatrics. A software overpressure monitor which vents if the pressure exceeds 300mmHg for adults and 200mmHg for pediatrics. If the hardware over-pressure circuit is tripped in normal operation then the unit must be turned off and back on to reset the system.

Inflation Source

  • This inflation source is capable of supplying sufficient air to bring a volume of 700cc’s to a pressure of 300mmHg in no more than 35 seconds. If the cuff is not inflated to the desired pressure within 60 seconds then the cuff is vented and a retry cycle is initiated.

Leak Rate

  • With the bleed valve closed, the maximum pressure drop shall be mmHg in 90 seconds measured with a 700cc volume at a differential pressure of 250mmHg.

Cuff Vent Rate

  • When the unit is vented, a volume of at least 700cc’s is reduce from a pressure of 250mmHg to a pressure of 20mmHg in a maximum of 14 seconds.

Temperature (Predictive)

  • Range: 90-110°F (32-43°C)
  • Display Resolution: 0.1°F (0.1°C)

SpO2

  • Range: 40-100% SpO2
  • Display Resolution: 1% SpO2
  • Display Update: Every 2 ±0.2 seconds
  • Calibration: Factory calibrated to functional saturation
  • Accuracy: ±2% from 70-100% SpO2; ±4% from 60-69% SpO2; Unspecified from 40-59% SpO2

Battery

  • Type: Sealed lead-acid
  • Number of Batteries: 1
  • Battery Voltage: 12 V DC nominal
  • Battery Capacity: 2.3 Amp-Hour
  • Battery Run Time: Accutorr Plus NIBP – 5 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes and recorder not in use; Accutorr Plus NIBP with Trend Screen – 3 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes and recorder not in use; Accutorr Plus NIBP with Trend Screen and SpO2 – 2 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes, continuous SpO2 measurement, and recorder not in use
  • Recharge Time: 8 hours maximum with unit in Standby
  • Cyclic Service Life: 150 cycles, 100% to 20% of capacity; 400 cycles, 100% to 40% of capacity

Real Time Clock

  • Resolution: 1 minute
  • Accuracy: ±1 minute/week
  • Display Format: 24 hours
  • Power: The real time clock maintains the time and date when the instrument is On or in the standby mode, connected to AC mains or running from internal battery for at least ten years from original assembly. The real-time clock will maintain time and date even if the instrument’s main battery is disconnected.